Skip to main content

Ethanol effects on Ga-radiolabelling efficacy and radiolysis in automated synthesis utilizing NaCl post-processing.

EJNMMI radiopharmacy and chemistry

Authors: Michael Meisenheimer, Stefan Kürpig, Markus Essler, Elisabeth Eppard

OBJECTIVE: Recent studies showed that ethanol in the reaction mixture improves radiolabelling with trivalent radiometals in terms of precursor amount, reaction time, reaction temperature and radiolysis. With regard to clinical application, this effect is of practical interest in radiopharmacy. The aim of this study was to evaluate whether the positive effect of ethanol can be exploited in automated systems utilizing NaCl-post processing.

METHODS: Gallium-68 was obtained from a 1.85 GBq Ge/Ga-generator. Radiolabelling was performed on an automated Ga-labelling cassette module. The standard labelling protocol was used without modifications. 0-40 vol% ethanol were added to the reaction mixture. Quality control was performed using radioHPLC and radioTLC.

RESULTS: Utilization of additional ethanol on an automated cassette module can be achieved by adding ethanol directly to the buffer solution without further modifications of the standard procedure. Radiolysis was decreased significantly as analysed by radioHPLC.

CONCLUSION: It was possible to combine the positive effects of ethanol on radiolabelling efficacy and radiolysis with the standard labelling procedure of an automated cassette module system. The whole process guarantees safe preparation of highly pure Ga-peptide for clinical application.

PMID: 31659535

Participating cluster members